Recall of INFANT RESUSCITATOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BAXTER CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    72987
  • Event Risk Class
    II
  • Event Initiated Date
    1996-03-22
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Maximum pressure relief valve setting is different to the 40 cm water described in the rd1000 operating manual could result in this safety feature not preventing delivery of greater inspiratory pressure to pat.

Device

  • Model / Serial
    Model Catalog: RD1000 (Lot serial: INCLUDES 93RD 94RD AND 95RD); Model Catalog: RD1000 (Lot serial: 92XXX0156 AND AFTER)
  • Product Description
    NEOPUFF INFANT RESUSCITATOR
  • Manufacturer

Manufacturer