  • Type of Event
    Recall
  • Event ID
    31557
  • Event Risk Class
    II
  • Event Initiated Date
    2011-10-31
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The movement of the warmer head can put stress on the head harness connector. prolonged exposure to this stress can lead to electrical arcing at the connector which may lead to discolouration occasionally charring of the electrical connector or it can lead to a loss of electrical power to the heating element.

Device

  • Model / Serial
    Model Catalog: IW931 (Lot serial: 010824 - 100930); Model Catalog: IW920 (Lot serial: 9891JEU00024 - 2001-98JEU00195); Model Catalog: IW930 (Lot serial: 9891JEU00024 - 2001-98JEU00195); Model Catalog: IW931 (Lot serial: 9891JEU00024 - 2001-98JEU00195); Model Catalog: IW932 (Lot serial: 9891JEU00024 - 2001-98JEU00195); Model Catalog: IW933 (Lot serial: 9891JEU00024 - 2001-98JEU00195); Model Catalog: IW910 (Lot serial: 9891JEU00024 - 2001-98JEU00195); Model Catalog: IW990 (Lot serial: 010824 - 100930); Model Catalog: IW980 (Lot serial: 010824 - 100930); Model Catalog: IW970 (Lot serial: 010824 - 100930); Model Catalog: IW960 (Lot serial: 010824 - 100930); Model Catalog: IW954 (Lot serial: 010824 - 100930); Model Catalog: IW953 (Lot serial: 010824 - 100930); Model Catalog: IW952 (Lot serial: 010824 - 100930); Model Catalog: IW951 (Lot serial: 010824 - 100930); Model Catalog: IW950 (Lot serial: 010824 - 100930); Model Catalog: IW934 (Lot serial: 010824 - 100930); Model Catalog: IW933 (Lot serial: 010824 - 100930
  • Product Description
    INFANT RADIANT WARMERS
  • Manufacturer