Recall of IMMULITE 2000 SYSTEM - TOTAL IGE ASSAY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27152
  • Event Risk Class
    III
  • Event Initiated Date
    2013-12-27
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens healthcare diagnostics has confirmed an overall average positive bias of 23% against the who 2nd irp 75/502 with the immulite systems total lge assays. the immulite total lge controls (ieca12) will not detect the bias.

Device

  • Model / Serial
    Model Catalog: L2KIE6 (Lot serial: LOTS 260 AND ABOVE); Model Catalog: L2KIE2 (Lot serial: LOTS 260 AND ABOVE)
  • Product Description
    IMMULITE 2000 SYSTEM-TOTAL IGE ASSAY