Recall of IMMULITE 2000 SYSTEM - GI-MA ASSAY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18668
  • Event Risk Class
    III
  • Event Initiated Date
    2015-02-13
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens healthcare diagnostics has confirmed a positive bias with the bio-rad lyphochek tumor marker plus control and the bio-rad liquichek tumor marker control when used with immulite 2000/immulite 2000 xpi gi-ma (ca19-9) assay kit lots 312 and 313. in addition a greater than expected change in patient sample recovery may be observed when moving from kit lot 311 to kit lots 312 and above. while recovery with patient samples for individual kit lots continues to meet siemens quality control release specification kit lot 311 was observed to perform at the lower end of the specification and kit lot 312 at the upper end of the specification resulting in a greater than expected difference between these two kit lots.

Device

  • Model / Serial
    Model Catalog: L2KGI2 (Lot serial: LOTS 312 AND 313)
  • Product Description
    IMMULITE 2000 SYSTEM - GI-MA ASSAY