Recall of IMMAGE IMMUNOCHEMISTRY SYSTEM RF RHEUMATOID FACTOR REAGENT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    66699
  • Event Risk Class
    II
  • Event Initiated Date
    2012-01-27
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Follow-up to recall 64398. beckman coulter has confirmed customer reports of qc shifts lot-to-lot variation and an increase in the precent of patient samples recovering >20 iu/ml using lots m012376 and m101865 of immage rheumatoid factor (rf) reagent. these lots of rf might produce false abnormal results.

Device

  • Model / Serial
    Model Catalog: 447070 (Lot serial: M101865); Model Catalog: 447070 (Lot serial: M012376)
  • Product Description
    IMMAGE Immunochemistry Systems Rheumatoid Factor (RF) Reagent
  • Manufacturer

Manufacturer