Recall of IL TEST (HEMOSIL) PT - FIBRINOGEN HS PLUS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by INSTRUMENTATION LABORATORY CO..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    31274
  • Event Risk Class
    II
  • Event Initiated Date
    2014-11-21
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Some vials of hemosil pt-fibrinogen hs plus pn 0008469810 exhibit a yellow/brownish color after reconstitution instead of the typical white to off-white color. these vials were observed to cause prolonged prothrombin (pt) clotting times. if an affected vial is used with a patient sample a significant shift in inr greater than 10% could occur with the potential to alter medical treatment.

Device

  • Model / Serial
    Model Catalog: 0008469810 (Lot serial: N0440009); Model Catalog: 0008469810 (Lot serial: N1122175); Model Catalog: 0008469810 (Lot serial: N0936197); Model Catalog: 0008469810 (Lot serial: N0735599); Model Catalog: 0008469810 (Lot serial: N0634953); Model Catalog: 0008469810 (Lot serial: N0132936); Model Catalog: 0008469810 (Lot serial: N0433932); Model Catalog: 0008469810 (Lot serial: N0349325); Model Catalog: 0008469810 (Lot serial: N0333382); Model Catalog: 0008469810 (Lot serial: N0249017); Model Catalog: 0008469810 (Lot serial: N0248652)
  • Product Description
    HemosIL PT-Fibrinogen HS PLUS