Recall of HP LCS IM HOLE LOCATOR INSTRUMENT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by JOHNSON & JOHNSON MEDICAL PRODUCTS A DIV. OF JOHNSON & JOHNSON INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    105168
  • Event Risk Class
    II
  • Event Initiated Date
    2015-01-29
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Depuy orthopaedics is issuing a device correction for all lots of hp lcs intramedullary (im) hole locator instruments because the japanese market has reported eight instances where femoral fractures occurred intra-operatively during primary knee replacement surgery while using the instrument. the company is not requesting the return of the device. this communication is to highlight and expand upon information from the appropriate surgical technique which states that the device may be lightly impacted for additional stability.