Recall of HOMECHOICE AUTOMATED PD SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BAXTER CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    71677
  • Event Risk Class
    II
  • Event Initiated Date
    2013-10-16
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    For dialysis patients with ascites using homechoice and homechoice pro devices with software version 10.4 because the device will attempt to drain the patient to empty during the initial drain there is an increased risk of hypotension and compromised hemodyamic status.

Device

  • Model / Serial
    Model Catalog: 5C4471 (Lot serial: >1000 NUMBERS CONTACT MFR); Model Catalog: 5C4471R (Lot serial: >1000 NUMBERS CONTACT MFR); Model Catalog: 5C8310 (Lot serial: >1000 NUMBERS CONTACT MFR); Model Catalog: 5C8310R (Lot serial: >1000 NUMBERS CONTACT MFR)
  • Product Description
    HOMECHOICE AUTOMATED PD SYSTEM
  • Manufacturer

Manufacturer