Recall of HOMECHOICE AUTOMATED PD SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BAXTER CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    145356
  • Event Risk Class
    II
  • Event Initiated Date
    2007-11-20
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    During a fill cycle if the system is powered down or a power failure occurs the system may not record the last 1 - 4 pump strokes infused into the patient when power is restored.

Device

  • Model / Serial
    Model Catalog: 5C4471R (Lot serial: MANUFACTURER); Model Catalog: 5C4471R (Lot serial: >100 LOT NUMBERS CONTACT); Model Catalog: 5C4471 (Lot serial: MANUFACTURER); Model Catalog: 5C4471 (Lot serial: >100 LOT NUMBERS CONTACT); Model Catalog: 5C8310 (Lot serial: >100 LOT NUMBERS CONTACT); Model Catalog: 5C8310P (Lot serial: >100 LOT NUMBERS CONTACT); Model Catalog: 5C8310R (Lot serial: >100 LOT NUMBERS CONTACT); Model Catalog: 5C8310 (Lot serial: MANUFACTURER); Model Catalog: 5C8310P (Lot serial: MANUFACTURER); Model Catalog: 5C8310R (Lot serial: MANUFACTURER)
  • Product Description
    HOME CHOICE AUTOMATED PD SYSTEM
  • Manufacturer

Manufacturer