Recall of HILL-ROM 100 LOW BED

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by HILL-ROM CANADA.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    100040
  • Event Risk Class
    II
  • Event Initiated Date
    2015-08-19
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The potential for a rare occurrence of a moderate injury has been identified. a laceration cut tear or abrasion to the skin on the back of the patient's lower leg could result in patients with frail skin if the leg is positioned at the opening between the thigh and foot sections or the thigh and head sections.

Device

  • Model / Serial
    Model Catalog: P3930A (Lot serial: >1000 numbers contact mfg.)
  • Product Description
    Hill-Rom 100 Low Bed
  • Manufacturer

Manufacturer