Recall of HER2 CISH PHARMDX KIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by DAKO DENMARK A/S.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    33982
  • Event Risk Class
    II
  • Event Initiated Date
    2016-06-16
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A wrong titration method was used when producing the probe mix for sk109 vial 3. this has resulted in the her2 dna probe concentration being approximately 4-5 times lower than usual for this product.

Device

  • Model / Serial
    Model Catalog: SK109 (Lot serial: 20031665)
  • Product Description
    HER2 CISH PHARMDX KIT;HER2 CISH PHarmdx kit
  • Manufacturer

Manufacturer