Recall of HEMOSIL RECOMBIPLASTIN

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by INSTRUMENTATION LABORATORY CO..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    34838
  • Event Risk Class
    III
  • Event Initiated Date
    2013-05-22
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    An on-board stability issue was identifed with the 8 ml size of hemosil recombiplastin 2g. the stability claim has been reduced to 3 days for all current and future product lots of this product.

Device

  • Model / Serial
    Model Catalog: 0020002950 (Lot serial: all lots)
  • Product Description
    HemosIL RecombiPlasTin 2G PN0020002950
  • Manufacturer

Manufacturer

  • Manufacturer Address
    BEDFORD
  • Manufacturer Parent Company (2017)
  • Source
    HC