Recall of HEMODIALYSIS 'Y' CONNECTORS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by CARDINAL HEALTH CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    104909
  • Event Risk Class
    II
  • Event Initiated Date
    2011-04-27
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Gambro has become aware of a possible defect on three lots of the gambro accessory s-660-c. the defect is a small hole at the level of the female luer connector that could potentially result in an intake of air into the bloodline. the air could subsequently reach the patient without passing through the "air blood" protective system of the dialysis machine. the small hole is not visible to the naked eye.

Device

  • Model / Serial
    Model Catalog: 101354-6431233 (Lot serial: 1035); Model Catalog: 101354-6431233 (Lot serial: 1036); Model Catalog: 101354-6431233 (Lot serial: 1037)
  • Product Description
    GAMBRO ACCESSORY S-660-C (HEMODIALYSIS Y CONNECTOR)
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Source
    HC