Recall of HEARTSTART MRX DEFIBRILLATOR/MONITOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    140722
  • Event Risk Class
    II
  • Event Initiated Date
    2014-02-25
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips has become aware that when used in hospital transport and pre-hospital (ems) environments the connection between the pads/paddles therapy cable (including pads cpr therapy cable external paddles cable and internal paddles cable/adapter) and the mrx therapy connection port could experience high levels of stress causing accelerated wear. wear of the connection could prevent the device from sensing that the pads/paddles therapy cable is connected. this wear also could cause the mrx to inappropriately identify the pads therapy cable external paddles or internal paddles. pads/paddles therapy cable and therapy connection port wear could pose a risk for patients and/or caregivers.

Device

  • Model / Serial
    Model Catalog: M3535A (Lot serial: ALL); Model Catalog: M3536A (Lot serial: ALL)
  • Product Description
    HEARTSTART MRX DEFIBRILLATOR/MONITOR
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC