Recall of HEARTSTART MRX DEFIBRILLATOR/MONITOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    65943
  • Event Risk Class
    II
  • Event Initiated Date
    2013-06-13
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The philips heartstart mrx monitor/defibrillator can deliver a non-synchronized shock in the synchronized cardioversion mode as follows: this rare condition can occur if the specific steps for delivering a synchronized shock in the heartstart mrx instructions for use are not followed. if the user rotates the therapy knob while simultaneously pressing the sync button then charges the mrx and presses the shock button the mrx could deliver a non-synchronized shock. the mrx will still display r wave markers as if normally operating in synchronized cardioversion mode. the "sync" label will still be seen on the display. when the mrx is charged and the shock button pressed the shock will be immediately delivered and may not be synchronized to the r wave. in addition the label affixed to the top of the mrx device is not consistent with the mrx instructions for use on synchronized cardioversion.

Device

  • Model / Serial
    Model Catalog: M3536A (Lot serial: >1000 numbers contact mfg); Model Catalog: M3535A (Lot serial: >1000 numbers contact mfg)
  • Product Description
    HEARTSTART MRX DEFIBRILLATOR/MONITOR
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC