Recall of HEARTSTART FR3 DEFIBRILLATOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    32268
  • Event Risk Class
    II
  • Event Initiated Date
    2014-02-21
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips has become aware that in a limited number fr3 aeds the printed circuit board (pcb) which contains the circuitry that supplies energy to the device could become separated from the electrical contact points that hold the board in place. in the event the pcb becomes separated from the electrical contact points the aed may not be ready to deliver a shock in the event of an emergency.

Device

  • Model / Serial
    Model Catalog: 861389 (Lot serial: ALL); Model Catalog: 861388 (Lot serial: ALL)
  • Product Description
    HEARTSTART FR3 DEFIBRILLATOR
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC