Recall of HEARTSINE SAMARITAN PAD WITH CPR ADVISOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by RESCUE 7 INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    105531
  • Event Risk Class
    II
  • Event Initiated Date
    2014-03-06
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The software in the samaritan pad 500p may miscalculate the cpr rate of compression per minute being administered to the patient. the rescuer may therefore be incorrectly advised by the device to "push slower" when in fact the cpr rate is at an acceptable level.

Device

  • Model / Serial
    Model Catalog: PAD 500P (Lot serial: 10B0010001 to 14B00461703)
  • Product Description
    SAMARITAN PAD SAM 500P
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC