Recall of GIRAFFE AND PANDA T-PIECE RESUSCITATION SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by MAQUET-DYNAMED INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    15595
  • Event Risk Class
    II
  • Event Initiated Date
    2009-09-30
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    1. moulding flash during manufacturing of t-piece at supplier caused out of specification aperture. 2. severity in risk analysis control document (rac) scored low leading to less strict supplier inspection and process control requirements.

Device

  • Model / Serial
    Model Catalog: M1091335 (Lot serial: NON AVAILABLE); Model Catalog: M1091365 (Lot serial: NON AVAILABLE); Model Catalog: M1091316 (Lot serial: NON AVAILABLE)
  • Product Description
    GIRAFFE & PANDA T-PIECE RESUS SYSTEM
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Source
    HC