Recall of GEMSTAR DOCKING STATION

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by HOSPIRA HEALTHCARE CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    108660
  • Event Risk Class
    II
  • Event Initiated Date
    2013-12-18
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Extra notes in the data
  • Reason
    The gemstar stations do not meet the hospira product specification for output voltage. under certain conditions the potential exists for the pump to fail to power up while connected to the docking station. moreover under a specific configuration the pump will display error code 11/003 indicating excessive input voltage from the external sources. this code will stop infusion if the error occurs while an infusion is in progress.

Device

  • Model / Serial
    Model Catalog: 13075-XX-YY (Lot serial: All serial number); Model Catalog: 13075 (Lot serial: All serial number)
  • Product Description
    Gemstar Docking Station
  • Manufacturer

Manufacturer

  • Manufacturer Address
    SAINT-LAURENT
  • Manufacturer Parent Company (2017)
  • Source
    HC