Recall of GEMSTAR BOLUS CORD

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by HOSPIRA HEALTHCARE CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    110315
  • Event Risk Class
    II
  • Event Initiated Date
    2012-03-14
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Extra notes in the data
  • Reason
    There is a potential of bolus delivery falilures when using the gemstar bolus cord. possible failures may be a result of cracked or damaged bolus cords which my result in an open or short circuit of the bolus cord. post-market incident reports indicate that the defects with these cords can lead to failure to deliver intermittent operation and unexpected delivery.

Device

  • Model / Serial
    Model Catalog: 13027 (Lot serial: All Serial Numbers)
  • Product Description
    Gemstar Bolus Cord
  • Manufacturer

Manufacturer

  • Manufacturer Address
    SAINT-LAURENT
  • Manufacturer Parent Company (2017)
  • Source
    HC