Recall of GEMSTAR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by HOSPIRA HEALTHCARE CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    28923
  • Event Risk Class
    II
  • Event Initiated Date
    2014-02-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Extra notes in the data
  • Reason
    In gemstar (phase 3) infusion pumps list # 13000 and 13150 the connection between the beeper subassembly and the pump may fail. the gemstar infusion pump will identify this failure during the "self-test" while powering up which will result in a beeper error (x100). this beeper error (x100) is a service alarm that places the pump in an inoperable mode and requires service before it can be returned to service.

Device

  • Model / Serial
    Model Catalog: 13000-XX-YY (Lot serial: All Serial Numbers); Model Catalog: 13150-XX-YY (Lot serial: All Serial Numbers)
  • Product Description
    GemStar Pumps