Recall of GALILEO ECHO INSTRUMENT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by IMMUCOR INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    114254
  • Event Risk Class
    III
  • Event Initiated Date
    2018-03-07
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Immucor inc. is issuing a medical device recall for galileo echo instrument. immucor has recently become aware of an issue within the echo software that will cause a software assertion error when loading reagents that have an expiry date greater than december 31 2019. the echo will not recognize products with an expiry date greater than december 31 2019.

Device

  • Model / Serial
    Model Catalog: 0087000 (Lot serial: Software Version 1.4.1.26.2)
  • Product Description
    Galileo Echo Instrument
  • Manufacturer

Manufacturer

  • Manufacturer Address
    NORCROSS
  • Manufacturer Parent Company (2017)
  • Source
    HC