Recall of FLEXOR CHECK-FLO PERFORMER INTRODUCER SET

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by COOK (CANADA) INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    22967
  • Event Risk Class
    II
  • Event Initiated Date
    2015-06-09
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    This recall is being initiated by cook medical due to reports received indicating the introducing dilator component extends beyond the introducer sheath further than intended and does not conform to the specifications. potential adverse events that may occur as a result of the excess dilator length include difficulty in accessing the hepatic vein requiring additional manipulation which could result in vessel injury.

Device

  • Model / Serial
    Model Catalog: KCFW-10.0-35-RB-RTPS-100 (Lot serial: 5805365); Model Catalog: KCFW-10.0-35-RB-RTPS-100 (Lot serial: 5770328); Model Catalog: KCFW-10.0-35-RB-RTPS-100 (Lot serial: 5770364)
  • Product Description
    Flexor RTPS Guiding Sheath
  • Manufacturer

Manufacturer