Recall of FABIUS TIRO ANESTHESIA SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by DRAEGER MEDICAL CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    30520
  • Event Risk Class
    I
  • Event Initiated Date
    2017-11-03
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Draeger has become aware of information suggesting that the use of the auxiliary common gas outlet (acgo) at anesthesia workstations of different manufacturers was not performed as described in the instructions for use of the relevant manufacturer and not according to pubiished safety guidelines (pre-use checklist) from various anesthesia societies.

Device

  • Model / Serial
    Model Catalog: 8606000 (Lot serial: ASDH-0190); Model Catalog: 8607300 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: 8607000 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Product Description
    FABIUS TIRO ANESTHESIA SYSTEM;FABIUS MRI;FABIUS GS PREMIUM SYSTEM
  • Manufacturer

Manufacturer