Recall of ESSENTA DR COMPACT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18528
  • Event Risk Class
    II
  • Event Initiated Date
    2012-07-05
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    1. during reliability testing it was determined that the u-arm could move down uncontrolled possibly from the breaking of a steel rope. 2. the screws for the sid chain wheel could be loose which could cause the detector carriage and/or the tube carriage to move down uncontrolled. 3. during in-house testing it was detected that the lifetime of the wire rope is significantly lower than the specified 10 years.

Device

  • Model / Serial
    Model Catalog: 9890 010 88201 (Lot serial: N/A)
  • Product Description
    ESSENTA DR COMPACT