Recall of ESSENTA DR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    151029
  • Event Risk Class
    II
  • Event Initiated Date
    2013-12-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips has become aware that if during system installation the screws holding the c-arm were mounted using thread locking compound (e.G. loctite) then the corresponding thread inserts inside the c-arm casting may loosen when the c-arm is dismantied. if after reassembly of the system the lossened thread inserts become detached from the casting the c-arm would no longer be securely held in position and could fall. the c-arm is most often dismantied and reassembled when the system is moved to another location.

Device

  • Model / Serial
    Model Catalog: 9890 010 87381 (Lot serial: 458280/SN10000031); Model Catalog: 9890 010 87381 (Lot serial: 433527/SN09000054); Model Catalog: 9890 010 87381 (Lot serial: 481201/SN11000021); Model Catalog: 9890 010 87381 (Lot serial: 423573/SN09000014); Model Catalog: 9890 010 87381 (Lot serial: 376657/SN08000030); Model Catalog: 9890 010 87381 (Lot serial: 396940/SN08000073); Model Catalog: 9890 010 87381 (Lot serial: 397186/SN08000042)
  • Product Description
    ESSENTA DR
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC