Recall of EP-4 CLINICAL STIMULATOR - TOUCH SCREEN

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BAYLIS MEDICAL COMPANY INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    32670
  • Event Risk Class
    I
  • Event Initiated Date
    2006-04-12
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Certain hardware and software components could lead to unanticipated device behaviors. (ex. initiation of a rapid stimulation train without command when in use in conjunction with an electrosurgical unit).

Device

  • Model / Serial
    Model Catalog: 09-1694-0000 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 09-1527-0000 (Lot serial: ALL SERIAL NUMBERS)
  • Product Description
    EP-4TM COMPUTERIZED STIMULATOR
  • Manufacturer

Manufacturer