Recall of ENGSTROM CARESTATION - MAIN UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by GENERAL ELECTRIC CANADA (OPERATING AS GE HEALTHCARE).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    51399
  • Event Risk Class
    II
  • Event Initiated Date
    2015-01-27
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Specific combinations of engstrom ventilator casters and bases have been identified that pose an increased likelihood of failures ofthe caster mounting hardware. a failure of the caster mounting hardware may result in the ventilator falling/tipping (overbalancing). a fall/tip of the engstrom ventilator can result in serious injury to patients or staff.

Device

  • Model / Serial
    Model Catalog: 1505-9000-000 (Lot serial: CBCP01475); Model Catalog: 1505-9000-000 (Lot serial: CBCP01415); Model Catalog: 1505-9000-000 (Lot serial: CBCP01452)
  • Product Description
    Engstrom carestation
  • Manufacturer

Manufacturer