Recall of ECLIPSE TPS V8.8

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by VARIAN MEDICAL SYSTEMS INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    85899
  • Event Risk Class
    II
  • Event Initiated Date
    2012-06-20
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A software anomaly has been identified with the eclipse treatment planning system where changing the prescription in a calculated vmat or rapid arc plan may lead to monitor units which do not reflect the new prescription. if the dose prescription of a vmat or rapidarc plan is changed after the initial dose volume optimization the monitor units and dose distribution may not reflect the intended prescription.

Device

  • Model / Serial
    Model Catalog: 100042024 (Lot serial: >10 Serial numbers.); Model Catalog: 100042024 (Lot serial: Contact manufacturer.)
  • Product Description
    Eclipse Treatment Planning System
  • Manufacturer

Manufacturer