Recall of DS CLIP APPLIERS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by AESCULAP AG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    107150
  • Event Risk Class
    II
  • Event Initiated Date
    2018-01-12
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    During review of the ifu it was detected additional missing contraindications and reprocessing instructions also required revision. there is no clinical data linked to the usage of the ds clips for the closure of the renal artery or renal vein during nephrectomies in living donor patients aesculap wants to avoid any potential patient risk by including this and other missing contraindications and also revising the cleaning and sterilization instructions.

Device

  • Model / Serial
    Model Catalog: PL817R (Lot serial: N/A); Model Catalog: PL806R (Lot serial: N/A); Model Catalog: PL462SU (Lot serial: N/A); Model Catalog: PL459SU (Lot serial: N/A); Model Catalog: PL465SU (Lot serial: N/A)
  • Product Description
    DS CLIP APPLIERS;CHALLENGER LIGATURE AND MARKING CLIPS
  • Manufacturer

Manufacturer