Recall of DRAEGER BREATHING BAG

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by DRAEGER MEDICAL CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    50663
  • Event Risk Class
    II
  • Event Initiated Date
    2014-03-18
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Draeger became aware of an incident where a user suffered an acute allergic reaction after using a draeger latex test lung. the user did not suffer permanent injury. it was discovered that the breathing bag that belongs to the test lung and other latex breathing bags are labelled "nr". although this is the standard abbreviation for natural rubber (latex) this labelling does not comply with the applicable international standards for medical devices.

Device

  • Model / Serial
    Model Catalog: 2165694 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: 2165953 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: 8403201 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Product Description
    REUSABLE LATEX BREATHING BAGS
  • Manufacturer

Manufacturer