Recall of DLP PRESSURE DISPLAY SET DISPOSABLE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by MEDTRONIC OF CANADA LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    26599
  • Event Risk Class
    III
  • Event Initiated Date
    2011-09-13
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The purpose of this field action is to inform customers of issues related to the directions for use (dfu) provided with the pressure display box and pressure display set. these issues are accuracy in the negative pressure range and length of use. medtronic has become aware that the pressure display box may not always accurately display readings. the negative pressure inaccuracy is relevant in applications involving "high vacuum" settings. these applications would be vacuum assisted venous drainage (vavd) and port access/mini sternotomy procedures. when used for an extended period (>24 hours) the positive pressure display may drift due to air leakage in the system.

Device

  • Model / Serial
    Model Catalog: 62006 (Lot serial: >100 numbers contact mfg); Model Catalog: 62000 (Lot serial: >100 numbers contact mfg); Model Catalog: 61006 (Lot serial: >100 numbers contact mfg); Model Catalog: 61000 (Lot serial: >100 numbers contact mfg); Model Catalog: 60000 (Lot serial: >10 numbers contact mfg); Model Catalog: 66000 (Lot serial: >10 numbers contact mfg); Model Catalog: 64000 (Lot serial: >10 numbers contact mfg); Model Catalog: 63000 (Lot serial: >10 numbers contact mfg)
  • Product Description
    DLP PRESSURE DISPLAY SET
  • Manufacturer

Manufacturer

  • Manufacturer Address
    BRAMPTON
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    HC