Recall of DIMENSION XPAND SYSTEM - CREATININE (CREA) ASSAY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    60979
  • Event Risk Class
    III
  • Event Initiated Date
    2015-07-21
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens has confirmed hemoglobin (hb) interference (greater than 10 percent) for dimension crea above 300 mg/dl hb at a creatinine concentration of 1.5 mg/dl and for dimension vista crea above 200 mg/dl hb at a creatinine concentration of 1.5 mg/dl. current claims for hb interference are less than 10 percent at these crea and hb concentrations.

Device

  • Model / Serial
    Model Catalog: DF33A (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: K1033 (Lot serial: >10 NUMBERS CONTACT MFR)
  • Product Description
    DIMENSION XPAND SYSTEM CREATININE(CREA)
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC