Recall of DIMENSION VISTA SYSTEM - IRON ASSAY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    53871
  • Event Risk Class
    III
  • Event Initiated Date
    2012-10-02
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens healthcare diagnostics has confirmed the potential for falsely elevated results with iron flex reagent cartridges when tests are processed from a well set contaminated with iron from environmental sources. all lots of iron flex reagent cartridges have the potential for this issue.

Device

  • Model / Serial
    Model Catalog: K3085 (Lot serial: > 10 NUMBERS CONTACT MFR)
  • Product Description
    IRON FLEX REAGENT CARTRIDGE