Recall of DIMENSION VISTA SYSTEM - CHEM CALIBRATORS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE DIAGNOSTICS INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    55666
  • Event Risk Class
    III
  • Event Initiated Date
    2010-02-25
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens healthcare diagnostics received customer complaints regarding dimension vista system chemistry 2 calibrator kc120. an alert message "calibrator insert missing" comes up when scanning the vial barcode for chem 2 cal level b lot 9gm001. an investigation of the barcode indicates that there are two additional digits in the barcode sequence.