Recall of DIMENSION VISTA SYSTEM - CHEM CALIBRATORS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    22700
  • Event Risk Class
    III
  • Event Initiated Date
    2009-09-09
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens has confirmed that this lot number may result in high recovery of quality control results show a positive bias of quality control results of up to 23% at uric acid concentrations of 3.5 mg/dl and 9.6 mg/dl.

Device

  • Model / Serial
    Model Catalog: KC110 (Lot serial: 9DM001)
  • Product Description
    DIMENSION VISTA SYSTEM-CHEM CALIBRATORS
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC