Recall of DIMENSION CLINICAL CHEMISTRY SYSTEM - IRON FLEX REAGENT CARTRIDGE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    99319
  • Event Risk Class
    III
  • Event Initiated Date
    2016-09-02
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens healthcare diagnostics has become aware of deferoxamine interference with dimension and dimension vista iron assays at a different concentration than listed in the instructions for use (ifu). deferoxamine is listed in the dimension and dimension vista ifu's as a non-interfering substance. the concentration that was used for interference testing (250 ng/dl or 3.8 nmol/l) is significantly below what would be expected in clinical practice. siemens healthcare diagnostics is providing an update to the limitation of procedure section for the dimension and dimension vista iron assays instructions for use (ifu) to indicate that: patients treated with metal-binding drugs (e.G. deferoxamine) may have depressed iron values as chelated iron may not properly react in the iron assay.

Device

  • Model / Serial
    Model Catalog: DF85 (Lot serial: All Lots); Model Catalog: K3085 (Lot serial: All Lots)
  • Product Description
    DIMENSION CLINICAL CHEMISTRY SYSTEM - IRON FLEX REAGENT CARTRIDGE;DIMENSION VISTA SYSTEM - IRON ASSAY;DIMENSION XPAND SYSTEM - IRON (IRON) ASSAY;DIMENSION EXL SYSTEM - IRON (IRON) ASSAY