Recall of DIMENSION CHEMISTRY SYSTEM - RXL MAX ANALYZER CLASS 3

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by DADE BEHRING CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    36078
  • Event Risk Class
    III
  • Event Initiated Date
    2006-12-20
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Manufacturer has identified an issue during an uncontrolled power down the instrument may lose track of the used aliquot segment well map and upon restart the instrument may reuse aliquot wells.

Device

  • Model / Serial
    Model Catalog: 420297.9731 (Lot serial: SOFTWARE VERSION 7.3.1)
  • Product Description
    SOFTWARE 7.3.1 WHEN USED WITH STREAMLAB
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Source
    HC