Recall of CYBERKNIFE ROBOTIC RADIOSURGERY SYSTEM - BASE CYBERKNIFE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ACCURAY INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    56322
  • Event Risk Class
    II
  • Event Initiated Date
    2009-07-08
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A healthcare facility reported their end-to-end targeting accuracy was out of specification. the position of one of the robotic manipulator axes differed from the position determined by the robotic controller. error 103 (karel e-stop detected) and error 342 (deviation in absolute position value dse - rdc) were displayed.

Device

  • Model / Serial
    Model Catalog: 028000 (Lot serial: C202)
  • Product Description
    CyberKnife Robotic Radiosurgery System
  • Manufacturer

Manufacturer

  • Manufacturer Address
    SUNNYVALE
  • Manufacturer Parent Company (2017)
  • Source
    HC