Recall of CXI SUPPORT CATHETER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by COOK (CANADA) INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    105405
  • Event Risk Class
    III
  • Event Initiated Date
    2015-10-27
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    This recall is being initiated by cook medical in response to six product complaints associated with reports of the 2.6fr curved catheter being mixed with the 4.0fr straight catheter and vice versa. investigation revealed the products were mixed during processing.

Device

  • Model / Serial
    Model Catalog: CXI-2.6-18-90-P-NS-ANG (Lot serial: 5505259)
  • Product Description
    CXI SUPPORT CATHETER
  • Manufacturer

Manufacturer