Recall of CURLIN 4000 PUMP KIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by B. BRAUN MEDICAL INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    26570
  • Event Risk Class
    I
  • Event Initiated Date
    2008-06-23
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Dropping or other physical abuse of the units can deform the pump chamber door resulting in unauthorized gravity/free flow of medications to occur resulting in over- medication of patients.

Device

  • Model / Serial
    Model Catalog: 350-1100 350-1200 (Lot serial: >10 #s contact manufacturer); Model Catalog: 350-1001 350-1000 (Lot serial: >10 #s contact manufacturer); Model Catalog: 360-1200 CUSTOM (Lot serial: >10 #s contact manufacturer); Model Catalog: 360-1301 360-1300 (Lot serial: >10 #s contact manufacturer); Model Catalog: 340-1001 340-1000 (Lot serial: ); Model Catalog: 340-1001 340-1000 (Lot serial: 102084); Model Catalog: 340-1001 340-1000 (Lot serial: 102086); Model Catalog: 340-1001 340-1000 (Lot serial: 102087); Model Catalog: 340-1001 340-1000 (Lot serial: 102083)
  • Product Description
    Curlin 4000 Pump kit 4000 CMS Kits
  • Manufacturer

Manufacturer