Recall of CRYOSTAT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by FISHER SCIENTIFIC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    142713
  • Event Risk Class
    II
  • Event Initiated Date
    2010-02-17
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Cryostat seriel 45847 may have been built using an incorrect cable with undersized diameter. it could potentially cause cable to overheat and to break. if instrument not grouned according to ifu a severe risk of electric shock may result.

Device

  • Model / Serial
    Model Catalog: 36 101 3493 (Lot serial: 45847)
  • Product Description
    hm550OVP cryostat
  • Manufacturer

Manufacturer