Recall of CRW SYSTEM - INTUBATION HEAD RING KIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by INTEGRA BURLINGTON MA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    149694
  • Event Risk Class
    III
  • Event Initiated Date
    2011-01-14
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Intubation hoops utilized in the manufacture of the hr-im intubation head ring assembly have been produced with excess material beyond the pivot drill hole. when pivoting the intubation hoop into the down position this excess material can come in contact with mating accessories making it impossible to (a) pivot the hoop into its final downward position or (b) prevent the accessories from being fully seated into the head ring assembly.

Device

  • Model / Serial
    Model Catalog: HRAIM (Lot serial: S/N: 0189463)
  • Product Description
    CRW SYSTEM - INTUBATION HEAD RING KIT
  • Manufacturer

Manufacturer