Recall of CONTINUUM TRILOGY IT ALLOFIT IT ACETABULAR SYSTEMS METASUL TAPER LINER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ZIMMER BIOMET CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    80265
  • Event Risk Class
    III
  • Event Initiated Date
    2016-01-22
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    In a follow-up correction zimmer biomet is removing from the field a subset (from zimmer gmbh) of highly polished implants that were packaged in a low density polyethylene (ldpe) bag with a potential to adhere to the implants. in september 2013 customers were asked to examine devices before use and return any showing signs of ldpe adhesion. at that time a new bag was introduced that resolves the issue and prevents the bag from adhering to the implants.

Device

  • Model / Serial
    Model Catalog: 00-8770-016-40 (Lot serial: >10 LOTS CONTACT MFR); Model Catalog: 00-8770-012-40 (Lot serial: >10 LOTS CONTACT MFR); Model Catalog: 00-8770-011-40 (Lot serial: >10 LOTS CONTACT MFR); Model Catalog: 00-8770-010-36 (Lot serial: >10 LOTS CONTACT MFR); Model Catalog: 00-8770-007-32 (Lot serial: >10 LOTS CONTACT MFR); Model Catalog: 00-8770-006-28 (Lot serial: >10 LOTS CONTACT MFR); Model Catalog: 01.04555.525 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 00-8770-040-02 (Lot serial: 25657381P); Model Catalog: 01.04555.505 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 01.04555.500 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 01.04555.485 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 01.04555.480 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 01.04555.460 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 01.04555.440 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 01.04555.420 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 01.04555.400 (Lot ser
  • Product Description
    CONTINUUMTRILOGYALLOFITMETASUL TAPER
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    HC