Recall of COMBI SET - HEMODIALYSIS BLOOD TUBING SET

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by FRESENIUS MEDICAL CARE CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    103243
  • Event Risk Class
    II
  • Event Initiated Date
    2013-06-28
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Based on recent complaints of significant blood loss (>100ml) and patient harm fresenius is undertaking to reinforce the importance of tightening all connections and providing additional detail on the best way to achieve a tight connection. this notice is specifically targetted to home programs where the end user is dialyzing in the home without the additional surveillance of a professional health care provider typically present in a clinical setting.

Device

  • Model / Serial
    Model Catalog: 03-2962-3 (Lot serial: >1000 NUMBERS CONTACT MFG); Model Catalog: 03-2622-3 (Lot serial: >1000 NUMBERS CONTACT MFG); Model Catalog: 03-2698-3 (Lot serial: >1000 NUMBERS CONTACT MFG); Model Catalog: 03-2695-9 (Lot serial: >1000 NUMBERS CONTACT MFG); Model Catalog: 03-2794-0 (Lot serial: >1000 NUMBERS CONTACT MFG); Model Catalog: 03-2696-7 (Lot serial: >1000 NUMBERS CONTACT MFG)
  • Product Description
    Combi Set - Hemodialysis blood tubing set
  • Manufacturer

Manufacturer