Recall of COBAS INTEGRA GLUCOSE HK

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ROCHE DIAGNOSTICS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    98403
  • Event Risk Class
    II
  • Event Initiated Date
    2001-11-14
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Erroneous results compared to the specif. of up to 40 mmol/l a reduced linearity of only 32 mmol/l was confirmed.Linearity cannot be reached reagent degrading faster than expected.

Device

  • Model / Serial
    Model Catalog: (Lot serial: LOT 624277); Model Catalog: (Lot serial: LIC 671); Model Catalog: (Lot serial: CAT 763020)
  • Product Description
    COBAS INTEGRA SYSTEM GLUCL
  • Manufacturer

Manufacturer

  • Manufacturer Address
    LAVAL
  • Manufacturer Parent Company (2017)
  • Source
    HC