Recall of CLOVERSNARE 4-LOOP VASCULAR RETRIEVER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by COOK (CANADA) INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    36355
  • Event Risk Class
    II
  • Event Initiated Date
    2014-07-18
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    We have received reports of instances where the snare loop has separated from the shaft of the device. the issue appears to be related to/caused when lateral forces are applied to the loop/shaft connection (e.G. in three reports separation occurred after attempts by a user to reshape the tip of the device). the issue was uncovered during a customer complaint investigation.

Device

  • Model / Serial
    Model Catalog: VRS-6.0-90 (Lot serial: 4319583); Model Catalog: VRS-6.0-90 (Lot serial: 4319577); Model Catalog: VRS-6.0-90 (Lot serial: 4319579); Model Catalog: VRS-6.0-90 (Lot serial: 4319581)
  • Product Description
    CLOVERSNARE 4-LOOP VASCULAR RETRIEVER
  • Manufacturer

Manufacturer