Recall of CLASSIC MAC LARYNGOSCOPE BLADE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by TRUDELL MEDICAL MARKETING LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    28501
  • Event Risk Class
    II
  • Event Initiated Date
    2011-01-20
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The greenlight ii laryngoscope handle with a top window battery cartridge may not function correctly. if the handle is subjected to a force perpendicular to the long axis of the handle a battery may shift in the cartridge. this could result in an intermittent light source or loss of light source.

Device

  • Model / Serial
    Model Catalog: 4558GSP (Lot serial: >10 numbers contact mfg)
  • Product Description
    GREENLIGHT II HANDLE AND CARTRIDGES
  • Manufacturer

Manufacturer