Recall of CARESCAPE R860

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by GENERAL ELECTRIC CANADA (OPERATING AS GE HEALTHCARE).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    78518
  • Event Risk Class
    II
  • Event Initiated Date
    2015-07-28
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The following issue is being addressed by ge healthcare file no. fmi 34066: the evair component manufacturer imtmedical ag is conducting a field corrective action. imtmedical has identified that potential friction of cables and tubing internal to the evair compressor may result in an inability for the compressor to output compressed air. there is a potential for temporary oversupply of oxygen to the patient as a result of this failure because the ge healthcare ventilators are designed to ventilate with 100% oxygen in the case of loss of input air supply pressure.