Recall of CADD ADMINISTRATION SET

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SMITHS MEDICAL CANADA LTD. (SIMS).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27754
  • Event Risk Class
    III
  • Event Initiated Date
    2017-03-15
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A drawing on the ifu incorrectly represents the atual filter size of the contents. where the drawing should indicate that the package contain an administration set with a 0.2 micron filter it is incorrectly illustrated as a 2.0 micron filter. the actual product in the package and the text based description on the ifu correctly conformed to the 0.2 micron filter specification for p/n 21-7339-24. it was the filter label on the drawing that was incorrect.

Device

  • Model / Serial
    Model Catalog: 21-7339-24 (Lot serial: 46x990); Model Catalog: 21-7339-24 (Lot serial: 46x896); Model Catalog: 21-7339-24 (Lot serial: 46x899); Model Catalog: 21-7339-24 (Lot serial: 46x829); Model Catalog: 21-7339-24 (Lot serial: 46x817); Model Catalog: 21-7339-24 (Lot serial: 46x831); Model Catalog: 21-7339-24 (Lot serial: 46x897); Model Catalog: 21-7339-24 (Lot serial: 46x1001); Model Catalog: 21-7339-24 (Lot serial: 46x828)
  • Product Description
    CADD Administration Set
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC